Model Number N/A |
Device Problem
Detachment of Device or Device Component (2907)
|
Patient Problem
No Patient Involvement (2645)
|
Event Type
malfunction
|
Manufacturer Narrative
|
(b)(4).Product has been received by zimmer biomet and the investigation is in process.Once the investigation has been completed, a follow-up mdr will be submitted.
|
|
Event Description
|
It was reported that the ball bearing fell out of the top of the persona p-ref sizer so the boom won't click in for accurate sizing.This happened during the wash cycle.
|
|
Event Description
|
No further event information available at the time of this report.
|
|
Manufacturer Narrative
|
Complaint sample was evaluated and the reported event was confirmed.Visual examination of the returned product confirms that the ball bearing was missing from the instrument.Also, the instrument shows discoloration and nicks which can be related to repeated use.Device history record (dhr) was reviewed and no discrepancies were found.A definitive root cause cannot be determined.If any further information is found which would change or alter any conclusions or information, a supplemental report will be filed accordingly.Zimmer biomet will continue to monitor for trends.
|
|
Search Alerts/Recalls
|