Model Number N/A
Device Problem Detachment of Device or Device Component (2907)
Patient Problem No Patient Involvement (2645)
Event Type  malfunction  
Manufacturer Narrative
(b)(4).Product has been received by zimmer biomet and the investigation is in process.Once the investigation has been completed, a follow-up mdr will be submitted.
 
Event Description
It was reported that the ball bearing fell out of the top of the persona p-ref sizer so the boom won't click in for accurate sizing.This happened during the wash cycle.
 
Event Description
No further event information available at the time of this report.
 
Manufacturer Narrative
Complaint sample was evaluated and the reported event was confirmed.Visual examination of the returned product confirms that the ball bearing was missing from the instrument.Also, the instrument shows discoloration and nicks which can be related to repeated use.Device history record (dhr) was reviewed and no discrepancies were found.A definitive root cause cannot be determined.If any further information is found which would change or alter any conclusions or information, a supplemental report will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
POSTERIOR REFERENCING SIZER
Type of Device
PROSTHESIS, KNEE
Manufacturer (Section D)
ZIMMER BIOMET, INC.
56 e. bell drive
warsaw IN 46582
MDR Report Key11078414
MDR Text Key223811601
Report Number0001822565-2020-04223
Device Sequence Number1
Product Code MBH
UDI-Device Identifier00889024232556
UDI-Public(01)00889024232556
Combination Product (y/n)N
PMA/PMN Number
K172524
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type distributor
Type of Report Initial,Followup
Report Date 02/17/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberN/A
Device Catalogue Number42509904000
Device Lot NumberZB6703400
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/23/2020
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 12/10/2020
Initial Date FDA Received12/28/2020
Supplement Dates Manufacturer Received02/12/2021
Supplement Dates FDA Received02/17/2021
Is This a Reprocessed and Reused Single-Use Device? No
Removal/Correction NumberN/A
Patient Sequence Number1
-
-