• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: OLYMPUS MEDICAL SYSTEMS CORP. HIGH FLOW INSUFFLATION UNIT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

OLYMPUS MEDICAL SYSTEMS CORP. HIGH FLOW INSUFFLATION UNIT Back to Search Results
Model Number UHI-4
Device Problem Loss of Power (1475)
Patient Problem No Patient Involvement (2645)
Event Type  malfunction  
Manufacturer Narrative
The suspect device was returned to olympus and evaluated.The reported problem was confirmed; upon powering on the device, the alarm began and all led's of the front panel indicator turned off.The cause of the problem was isolated to a faulty main board.
 
Event Description
A user facility reported to olympus that the device shut off and started alarming.There was no patient injury or harm, associated with the problem, reported to olympus.
 
Manufacturer Narrative
This report is being supplemented to provide additional information based on the legal manufacturer's final investigation.A review of the device history record (dhr) found no deviations that could have caused or contributed to the reported issue.Based on the results of the investigation, the probable cause is likely the pressure sensor on the main board failed, causing the phenomenon.The failure of the pressure sensor was caused by the following factors.The amount of use and age of the device (over 7 years).Process abnormalities such as: corrosion of the pressure sensor, foreign matter adhering to the wiring inside the pressure sensor.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HIGH FLOW INSUFFLATION UNIT
Type of Device
HIGH FLOW INSUFFLATION UNIT
Manufacturer (Section D)
OLYMPUS MEDICAL SYSTEMS CORP.
2951 ishikawa-cho
hachioji-shi, tokyo-to 192-8 507
JA  192-8507
MDR Report Key11080711
MDR Text Key227355096
Report Number8010047-2020-11045
Device Sequence Number1
Product Code HIF
UDI-Device Identifier04953170324147
UDI-Public04953170324147
Combination Product (y/n)N
PMA/PMN Number
K122180
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type user facility
Type of Report Initial,Followup
Report Date 02/23/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberUHI-4
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/12/2020
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 12/01/2020
Initial Date FDA Received12/28/2020
Supplement Dates Manufacturer Received02/01/2021
Supplement Dates FDA Received02/23/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-