This report is being supplemented to provide additional information based on the legal manufacturer's final investigation.A review of the device history record (dhr) found no deviations that could have caused or contributed to the reported issue.Based on the results of the investigation, the probable cause is likely the pressure sensor on the main board failed, causing the phenomenon.The failure of the pressure sensor was caused by the following factors.The amount of use and age of the device (over 7 years).Process abnormalities such as: corrosion of the pressure sensor, foreign matter adhering to the wiring inside the pressure sensor.
|