Model Number 1458Q/86 |
Device Problems
Difficult to Insert (1316); Separation Failure (2547)
|
Patient Problems
No Consequences Or Impact To Patient (2199); No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 12/07/2020 |
Event Type
malfunction
|
Manufacturer Narrative
|
This product is registered as a combination product.The results/method and conclusion codes along with investigation results will be provided in the final report.
|
|
Event Description
|
It was reported that patient presented for a lead implant surgery.During procedure, the wire was getting stuck to the lead, not being able to pass.The lead was not used and it was replaced by a new one.The patient was stable.
|
|
Manufacturer Narrative
|
The reported event of a guidewire insertion issue was confirmed.As received, a complete lead was returned for analysis.The cause of the reported event was isolated to the damaged guidewire consistent with procedural damage.
|
|