• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ST. JUDE MEDICAL, INC.(CRM-SYLMAR) QUARTET; PERMANENT PACEMAKER ELECTRODE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ST. JUDE MEDICAL, INC.(CRM-SYLMAR) QUARTET; PERMANENT PACEMAKER ELECTRODE Back to Search Results
Model Number 1458Q/86
Device Problems Difficult to Insert (1316); Separation Failure (2547)
Patient Problems No Consequences Or Impact To Patient (2199); No Clinical Signs, Symptoms or Conditions (4582)
Event Date 12/07/2020
Event Type  malfunction  
Manufacturer Narrative
This product is registered as a combination product.The results/method and conclusion codes along with investigation results will be provided in the final report.
 
Event Description
It was reported that patient presented for a lead implant surgery.During procedure, the wire was getting stuck to the lead, not being able to pass.The lead was not used and it was replaced by a new one.The patient was stable.
 
Manufacturer Narrative
The reported event of a guidewire insertion issue was confirmed.As received, a complete lead was returned for analysis.The cause of the reported event was isolated to the damaged guidewire consistent with procedural damage.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
QUARTET
Type of Device
PERMANENT PACEMAKER ELECTRODE
Manufacturer (Section D)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
MDR Report Key11085307
MDR Text Key224027270
Report Number2017865-2020-23514
Device Sequence Number1
Product Code NIK
UDI-Device Identifier05414734503198
UDI-Public05414734503198
Combination Product (y/n)N
PMA/PMN Number
P030054
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Type of Report Initial,Followup
Report Date 03/22/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/28/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date09/30/2023
Device Model Number1458Q/86
Device Catalogue Number1458Q-86
Device Lot NumberA000102516
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/09/2020
Was the Report Sent to FDA? No
Date Manufacturer Received03/19/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Age56 YR
-
-