Catalog Number 5111-00300-060
Device Problem Device Dislodged or Dislocated (2923)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 11/23/2020
Event Type  malfunction  
Manufacturer Narrative
(b)(6).The device was received for evaluation.During visual investigation, it was noted that the returned product showed signs of use (dried blood) which is consistent with the reporter¿s statement that the alleged event occurred during use.The reported condition was verified.However, the cause of the condition could not be determined because the returned coupler rings had been successfully approximated (brought together aligned and tightly closed) and were no longer attached in the ¿u port) of the coupler jaw assembly.A device history review revealed no issues that could have caused or contributed to the reported issue.Should additional relevant information become available, a supplemental report will be submitted.
 
Event Description
It was reported that the ring of a 3.0mm coupler was not fully fixed on the u port and was unable to be pushed out of the anastomotic instrument.It was further reported that this issue was identified during patient use.The surgeon used another coupler to complete the case.There was no report of patient injury or medical intervention associated with this event.It was further reported, the ¿patient was doing well¿.No additional information is available.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
COUPLER
Type of Device
DEVICE, ANASTOMOTIC, MICROVASCULAR
Manufacturer (Section D)
BAXTER HEALTHCARE CORPORATION
deerfield IL
Manufacturer (Section G)
SYNOVIS SURGICAL INNOVATIONS
2575 university ave. w
saint paul MN 55114
Manufacturer Contact
25212 w. illinois route 120
round lake, IL 60073
2242702068
MDR Report Key11089228
MDR Text Key224755629
Report Number1416980-2020-08151
Device Sequence Number1
Product Code MVR
Combination Product (y/n)Y
Reporter Country CodeTW
PMA/PMN Number
K861985
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,distri
Reporter Occupation Other
Type of Report Initial
Report Date 12/29/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/29/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number5111-00300-060
Device Lot NumberSP20G071462251
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/04/2020
Date Manufacturer Received12/03/2020
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-