Device evaluated by mfr.: promus elite us mr 32 x 2.50mm stent delivery system was returned for analysis.A visual examination of the stent found proximal stent damage with the struts pulled in a distal direction.The undamaged crimped stent outer diameter was measured and the result is maximum crimped stent profile measurement.The balloon cones were reviewed, and no issues were noted.The balloon wings were tightly wrapped and evenly folded and were not subjected to positive pressure.A visual and microscopic examination of the bumper tip found no issues.A visual and tactile examination of the hypotube found no issues.A visual and tactile examination of the outer and mid-shaft section and a visual examination of the inner lumen found bunching/accordioning of the inner shaft polymer extrusion 138.2cm distal to the distal end of strain relief.An attempt was made to load test guidewire however when the tip of the guidewire reached the damage in the inner shaft polymer extrusion it could not advance further.No other issues were identified during the product analysis.
|