Device manufacture dates: battery: 11/15/2018, charger: 03/25/2011.Device evaluation of battery pack sn (b)(4) has been completed.The reported problem (battery faults) was confirmed.As received, the battery pack was unable to power a test monitor and charge.Upon evaluation, the nickel busbar tabs at the p2 & p4 pads were loose.The cause of the non-functional battery pack is the loose busbar.The root cause of the loose busbar cannot be positively identified.No adverse event resulted from the defective battery.Device evaluation of battery charger/modem sn (b)(4) has been completed.The reported problem (charger problem) has been confirmed.Upon investigation the battery charger/modem was unable to recharge a battery pack.The cause for the failure was contamination on the battery pca.The root cause for the contamination was ingress of an unknown liquid.Lifevest patient training materials have been updated as a reminder not to expose the lifevest electronic components to liquids.No adverse event resulted from the defective battery charger.
|