ZOLL MANUFACTURING CORPORATION LIFEVEST WCD 4000 SYSTEM; WEARABLE CARDIOVERTER DEFIBRILLATOR Back to Search Results
Model Number WCD 4000
Device Problems Battery Problem (2885); Therapeutic or Diagnostic Output Failure (3023)
Patient Problem Insufficient Information (4580)
Event Date 11/06/2020
Event Type  malfunction  
Manufacturer Narrative
Device manufacture dates: battery: 11/15/2018, charger: 03/25/2011.Device evaluation of battery pack sn (b)(4) has been completed.The reported problem (battery faults) was confirmed.As received, the battery pack was unable to power a test monitor and charge.Upon evaluation, the nickel busbar tabs at the p2 & p4 pads were loose.The cause of the non-functional battery pack is the loose busbar.The root cause of the loose busbar cannot be positively identified.No adverse event resulted from the defective battery.Device evaluation of battery charger/modem sn (b)(4) has been completed.The reported problem (charger problem) has been confirmed.Upon investigation the battery charger/modem was unable to recharge a battery pack.The cause for the failure was contamination on the battery pca.The root cause for the contamination was ingress of an unknown liquid.Lifevest patient training materials have been updated as a reminder not to expose the lifevest electronic components to liquids.No adverse event resulted from the defective battery charger.
 
Event Description
A us distributor reported that a patient was receiving battery fault and had a charger problem.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LIFEVEST WCD 4000 SYSTEM
Type of Device
WEARABLE CARDIOVERTER DEFIBRILLATOR
Manufacturer (Section D)
ZOLL MANUFACTURING CORPORATION
121 gamma drive
pittsburgh PA 15238 3495
Manufacturer (Section G)
ZOLL MANUFACTURING CORPORATION
121 gamma drive
pittsburgh PA 15238 3495
Manufacturer Contact
matthew saich
121 gamma drive
pittsburgh, PA 15238-3495
4129683333
MDR Report Key11091490
MDR Text Key224346037
Report Number3008642652-2020-11549
Device Sequence Number1
Product Code MVK
UDI-Device Identifier00855778005036
UDI-Public00855778005036
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P010030
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type distributor
Reporter Occupation Other
Type of Report Initial
Report Date 12/29/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/29/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Model NumberWCD 4000
Device Catalogue Number10A0989
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/01/2020
Date Manufacturer Received12/10/2020
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured11/15/2018
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-