BD RAPIS DIAGNOSTICS (SUZHOU) CO. LTD. BD RAPID DETECTION OF SARS-COV-2 VERITOR; CORONAVIRUS ANTIGEN DETECTION SYSTEM Back to Search Results
Catalog Number 256082
Device Problem Incorrect, Inadequate or Imprecise Result or Readings (1535)
Patient Problems No Consequences Or Impact To Patient (2199); No Clinical Signs, Symptoms or Conditions (4582)
Event Date 12/03/2020
Event Type  malfunction  
Manufacturer Narrative
A device evaluation and/or device history review is anticipated, but is not complete.Upon completion, a supplemental report will be filed.(b)(4).
 
Event Description
It was reported while testing for sars cov-2 a false positive result was obtained.A repeat test was performed 1-2 weeks later using pcr and the result was negative.The customer stated there was no patient impact.Eua#: (b)(4).
 
Manufacturer Narrative
H.6.Investigation: bd has received several customer complaints for false positive results, when using bd sars-cov-2 reagents for bd veritor¿ system.The current investigation concerns multiple lots of bd sars-cov-2 reagents for bd veritor¿ system.Bd takes a systematic approach to investigating false positive complaints that are received.This approach involves review of manufacturing batch history records for batch number 0242768, testing of retention samples of batch number 0242768, and testing of customer returned samples if applicable.The investigation did not find a root cause for the false positive results that were observed.Bd cannot confirm the complaint based on the investigation that was performed.A corrective and preventive action (capa#1878253) is already initiated to investigate the root cause and some mitigation actions are already being addressed.H3 other text : see h.10.
 
Event Description
It was reported while testing for sars cov-2 a false positive result was obtained.A repeat test was performed 1-2 weeks later using pcr and the result was negative.The customer stated there was no patient impact.Eua#: eua(b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BD RAPID DETECTION OF SARS-COV-2 VERITOR
Type of Device
CORONAVIRUS ANTIGEN DETECTION SYSTEM
Manufacturer (Section D)
BD RAPIS DIAGNOSTICS (SUZHOU) CO. LTD.
9 rui pu road export zone b
suzhou
CH 
MDR Report Key11091606
MDR Text Key224862745
Report Number3006948883-2020-01060
Device Sequence Number1
Product Code QKP
Combination Product (y/n)N
PMA/PMN Number
NA
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type other,user facility
Type of Report Initial,Followup
Report Date 04/29/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Expiration Date02/01/2021
Device Catalogue Number256082
Device Lot Number0242768
Was Device Available for Evaluation? No
Initial Date Manufacturer Received 12/03/2020
Initial Date FDA Received12/29/2020
Supplement Dates Manufacturer Received04/29/2021
Supplement Dates FDA Received05/07/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-