Model Number 9-ASD-032
Device Problem Material Deformation (2976)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 12/03/2020
Event Type  malfunction  
Manufacturer Narrative
The reported event of deformation upon deployment could not be confirmed.The investigation confirmed the device met visual and functional specifications when analyzed at abbott.The device history record was reviewed to ensure that each manufacturing and inspection operation was performed and the product met all specifications.The cause of the reported incident could not be conclusively determined.
 
Event Description
On (b)(6) 2020, a 32 mm amplatzer septal occluder was selected for implant.While deploying the device, a cobra deformation was noted.The device was retracted, removed, then replaced with a new 32 mm amplatzer septal occluder to resolve the event.The procedure was not extended longer than is clinically significant and the patient remained hemodynamically stable throughout.The patient is currently stable with no adverse events.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AMPLATZER SEPTAL OCCLUDER
Type of Device
TRANSCATHETER SEPTAL OCCLUDER
Manufacturer (Section D)
AGA MEDICAL CORPORATION
5050 nathan lane north
plymouth MN 55442
Manufacturer (Section G)
AGA MEDICAL CORPORATION
5050 nathan lane north
plymouth MN 55442
Manufacturer Contact
pamela yip
5050 nathan lane n
plymouth, MN 55442
6517565400
MDR Report Key11092320
MDR Text Key225816400
Report Number2135147-2020-00557
Device Sequence Number1
Product Code MLV
UDI-Device Identifier00811806010236
UDI-Public00811806010236
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P000039
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Physician
Remedial Action Replace
Type of Report Initial
Report Date 12/29/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/29/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number9-ASD-032
Device Catalogue Number9-ASD-032
Device Lot Number7374381
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/15/2020
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received12/03/2020
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured02/03/2020
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Age34 YR
-
-