Model Number HL-90
Device Problems Pumping Problem (3016); Noise, Audible (3273)
Patient Problem No Information (3190)
Event Type  malfunction  
Manufacturer Narrative
One level 1 hotline low flow system was received with the printed circuit board (pcb), power switch and retainer needing an upgrade.The pump was very noisy, the enclosure was cracked at the drain fitting causing a leak, both covers were cracked, the heater was defective, he microswitch was broken and the line cord was worn.The water tank was filled, the temperature check was attached, the line cord was plugged in and the device was turned on.The reported issue was able to be duplicated.Normal wear and tear caused the noisy pump.The pump was replaced to resolve the primary issue.
 
Event Description
Information was received indicating that a smiths medical level 1 hotline low flow system was making loud noises and the pump was failing.There was no reported adverse event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LEVEL 1 HOTLINE LOW FLOW SYSTEMS
Type of Device
WARMER, THERMAL, INFUSION FLUID
Manufacturer (Section D)
ST PAUL
1265 grey fox rd.
st. paul MN 55112
Manufacturer (Section G)
ST PAUL
1265 grey fox rd.
st. paul MN 55112
Manufacturer Contact
david halverson
6000 nathan lane north
minneapolis, MN 55442
MDR Report Key11094828
MDR Text Key224296153
Report Number3012307300-2020-12848
Device Sequence Number1
Product Code LGZ
UDI-Device Identifier50695085820909
UDI-Public50695085820909
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K911383
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type user facility
Reporter Occupation Biomedical Engineer
Type of Report Initial
Report Date 12/30/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/30/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberHL-90
Device Catalogue NumberHL-90
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/07/2020
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received12/03/2020
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured01/22/2005
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-