SMITH & NEPHEW, INC. SPIDER2 24V BATTERY CHARGER; BATTERY, REPLACEMENT, RECHARGEABLE Back to Search Results
Catalog Number 72203840
Device Problem Electrical /Electronic Property Problem (1198)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 12/08/2020
Event Type  malfunction  
Manufacturer Narrative
(b)(6).
 
Event Description
It was reported that two battery chargers did not work to charge up the spider 2 battery.Incident occurred while setting-up for the procedure.It is unknown if there was back-up device available and if a delay occurred.All available information has been disclosed.If additional information should become available, a supplemental report will be submitted accordingly.
 
Manufacturer Narrative
The reported device, intended for use in treatment, was received for evaluation.There was no relationship found between the returned device and the reported incident.A visual inspection was performed on the product and no issue was observed.A functional evaluation did not reveal any problems.The charger's indicator diode turned red indicating a fast charge was initiated.The charger fully charged the test battery.No product identification information was provided and thus a manufacturing record review could not be conducted.A complaint history review concluded this was a repeat issue.The reported failure was not verified and the root cause could not be determined since the reported malfunction could not be duplicated during the product evaluation process.Please refer to the instructions for use for recommendations on proper use of the device and potential troubleshooting methods to prevent future reoccurrence of the reported event.No containment or corrective actions are recommended at this time.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SPIDER2 24V BATTERY CHARGER
Type of Device
BATTERY, REPLACEMENT, RECHARGEABLE
Manufacturer (Section D)
SMITH & NEPHEW, INC.
76 s. meridian ave.
oklahoma city OK 73107
MDR Report Key11097492
MDR Text Key224585012
Report Number1643264-2020-02248
Device Sequence Number1
Product Code MOQ
Combination Product (y/n)N
PMA/PMN Number
EXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Type of Report Initial,Followup
Report Date 03/02/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/30/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number72203840
Was Device Available for Evaluation? No
Date Manufacturer Received02/05/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-