SMITH & NEPHEW, INC. R3 OFFSET IMPACTOR TIP; PROSTHESIS,KNEE,FEMOROTIBIAL,SEMI-CONSTRAINED,CEMENTED,METAL/POLY
|
Back to Search Results |
|
Model Number 71363077 |
Device Problems
Fracture (1260); Device Damaged Prior to Use (2284)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 11/10/2020 |
Event Type
malfunction
|
Manufacturer Narrative
|
The associated device, intended for use in treatment, was returned and evaluated.A visual inspection of the returned device confirmed the stated failure mode.The device is fractured.The piece was not returned, rendering the device inoperable. the device shows signs of extensive use.At this time, we do not have reason to suspect that the product failed to meet any product specifications at the time of manufacture.This device is a reusable instrument that can be exposed to numerous surgeries.Damage from prolonged use, misuse or rough handling are likely probable causes of the reported event.We recommend that all reusable instruments be routinely inspected for wear and damage and replaced as necessary.Based on this investigation, the need for corrective action is not indicated.Should additional information be received, the complaint will be reopened.No further investigation is warranted for this complaint; however, we will continue to monitor for future complaints and investigate as necessary. we consider this investigation closed.(b)(4).
|
|
Event Description
|
It was reported that the tip does not stay on impactor tip.No case involved.All available information has been disclosed.If additional information should become available, a supplemental report will be submitted accordingly.
|
|