• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ZOLL MANUFACTURING CORPORATION LIFEVEST WCD 4000 SYSTEM; WEARABLE CARDIOVERTER DEFIBRILLATOR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ZOLL MANUFACTURING CORPORATION LIFEVEST WCD 4000 SYSTEM; WEARABLE CARDIOVERTER DEFIBRILLATOR Back to Search Results
Model Number WCD 4000
Device Problem Failure of Device to Self-Test (2937)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 11/23/2020
Event Type  malfunction  
Manufacturer Narrative
Device evaluation of electrode belt sn (b)(4) has been completed.The reported problem (cannot run detect or treat) has been confirmed.Upon investigation the trunk cable connector was glued to the electrode belt receptacle of a monitor.The root cause for the damaged connector was physical abuse.No adverse event resulted from the defective belt.
 
Event Description
A us distributor reported that an electrode belt was unable to undergo incoming functional testing.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LIFEVEST WCD 4000 SYSTEM
Type of Device
WEARABLE CARDIOVERTER DEFIBRILLATOR
Manufacturer (Section D)
ZOLL MANUFACTURING CORPORATION
121 gamma drive
pittsburgh PA 15238 3495
Manufacturer (Section G)
ZOLL MANUFACTURING CORPORATION
121 gamma drive
pittsburgh PA 15238 3495
Manufacturer Contact
matthew saich
121 gamma drive
pittsburgh, PA 15238-3495
4129683333
MDR Report Key11097979
MDR Text Key225279969
Report Number3008642652-2020-11467
Device Sequence Number1
Product Code MVK
UDI-Device Identifier00855778005135
UDI-Public00855778005135
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P010030
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type distributor
Reporter Occupation Other
Type of Report Initial
Report Date 12/22/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/30/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Model NumberWCD 4000
Device Catalogue Number10A1025
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/23/2020
Date Manufacturer Received12/15/2020
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured02/05/2016
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-