It was reported that the customer was unable to clear the bubble alarm till the cardiohelp was restarted.Further a venous probe issue was reported.A picture showing the issue was provided.A getinge field service technician was sent for investigation on 2020-12-23, he could not reproduce the failure.He performed a full maintenance and all tests were passed successfully.According to the technician the venous probe case was reconnected by the customer before the maintenance.The review of the complaint was performed by getinge life cycle engineering on 2021-02-22.It was reported that on receipt of the bubble alarm, the circuit was clamped and the alarm was reset.No information if the circuit was checked for air or if air was removed.Further, the flow was set to zero after clamping.This procedure could have been caused the error "wrong target lpm/rpm" leading to a pump stop.Further it was reported that after that the hls module was disconnected and after use of the emergency drive it was reattached to the cardiohelp.Multiple error messages were reported most probable the error "wrong target lpm/rpm" followed by "pump disposable error - stop" was displayed.This error is triggered if there is a deviation between set target lpm/rpm and actual lpm/rpm e.G.If the disposable is disconnected without putting the lpm/rpm to zero.Setting target lpm/rpm to zero and then again to the desired values clears this error.In this case the whole cardiohelp unit was restarted.(refer instruction for use cardiohelp system v1.15, chapter 9.4 technical alarms) based on the results the reported failure "bubble unable to clear" could not be confirmed.The most probable root cause could be a handling error (e.G.Inexperienced user).The sales and service unit was advised to make the customer aware of the mentioned instruction for use.Based on the results the reported failure ¿venous probe issue¿ could be confirmed as the provided picture showed the open venous probe case.The most probable root cause is a mechanical impact/ hit or wear of the venous probe case (in use since 2014).The occurrence rate was calculated for the reported failure and it was determined that this is not a systemic issue.Therefore, no remedial action is required.The occurrence rate related to the reported issue is currently being monitored as part of maquet cardiopulmonary¿ s trending program and additional investigations or corrections will be implemented in case of adverse trending.
|