Model Number CI-1600-04 |
Device Problems
Mechanical Problem (1384); Output below Specifications (3004); Insufficient Information (3190)
|
Patient Problems
Insufficient Information (4580); Appropriate Clinical Signs, Symptoms, Conditions Term / Code Not Available (4581)
|
Event Date 09/24/2020 |
Event Type
malfunction
|
Event Description
|
The recipient reportedly experienced decreased performance.Programming adjustments were made, however, the issue did not resolve.The recipient's device was explanted.The recipient was reimplanted with another cochlear device.
|
|
Manufacturer Narrative
|
Additional information: section d.9.Advanced bionics considers the investigation into this reportable event as closed.Legal proceedings have prohibited the testing and failure analysis of the explanted device.As a result, no conclusion can be drawn at this time.If the legal proceedings allow for the analysis to be completed, the issue will be re-opened and the results of the analysis will be reported.This is the final report.Disclaimer: advanced bionics does not intend that this report be any admission of liability, fault or product defect.
|
|
Manufacturer Narrative
|
Correction: section h.6.Advanced bionics considers the investigation into this reportable event as closed.Advanced bionics received consent to analyze on (b)(6) 2022.The external visual inspection revealed the electrode was severed prior to receipt and silicone damage was observed on the top cover of the device.This is believed to have occurred during revision surgery.The device passed photographic imaging inspection.System lock was verified.The condition of the electrode prevented an electrical test from being performed.The device passed some of the electrical tests performed.The device passed the mechanical tests performed.The failure of this device is attributed to an electrode short in the electrode pocket.A corrective action was implemented.This version of the hires ultra device is no longer distributed.This is the final report.Disclaimer: advanced bionics does not intend that this report be any admission of liability, fault or product defect.
|
|
Search Alerts/Recalls
|