• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION DYNAGEN CRT-D; IMPLANTABLE DEVICE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION DYNAGEN CRT-D; IMPLANTABLE DEVICE Back to Search Results
Model Number G151
Device Problems High impedance (1291); Failure to Read Input Signal (1581); Connection Problem (2900); High Capture Threshold (3266)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 09/22/2020
Event Type  malfunction  
Manufacturer Narrative
This product has not been returned to boston scientific, and as a result, laboratory analysis could not be conducted.The associated investigation determined that this device exhibited intermittent impedance measurements with no conclusive evidence of a malfunction or inadequate lead-to-device connection; please refer to the description for more information regarding the specific circumstances of this event.Investigation has determined that this type of event is likely the result of an intermittent high impedance condition associated with the device spring contact and lead terminal ring.A design enhancement was implemented in 2020 to stabilize the electrical connection between the spring contact and the lead terminal.
 
Event Description
It was reported that this cardiac resynchronization therapy defibrillator (crt-d) and another manufacturer's lead exhibited high thresholds and increased pace impedance measurements.It was also reported that this device exhibited out of range intrinsic amplitude measurements.No adverse patient effects were reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DYNAGEN CRT-D
Type of Device
IMPLANTABLE DEVICE
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
4100 hamline avenue north
saint paul MN 55112
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
cashel road
clonmel
EI  
Manufacturer Contact
timothy degroot
4100 hamline avenue north
saint paul, MN 55112
6515826168
MDR Report Key11104983
MDR Text Key224898515
Report Number2124215-2020-26196
Device Sequence Number1
Product Code NIK
UDI-Device Identifier00802526534638
UDI-Public00802526534638
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P010012/S341
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 01/02/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Expiration Date09/13/2019
Device Model NumberG151
Device Catalogue NumberG151
Device Lot Number124248
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 11/22/2020
Initial Date FDA Received01/02/2021
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured10/04/2017
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Age64 YR
-
-