Model Number CI-1600-04 |
Device Problems
Mechanical Problem (1384); Output Problem (3005); Noise, Audible (3273)
|
Patient Problem
Insufficient Information (4580)
|
Event Date 03/11/2021 |
Event Type
malfunction
|
Event Description
|
The recipient is reportedly experiencing decreased performance and loudness growth issues.Programming adjustments were made, however, the issue did not resolve.Revision surgery is scheduled.
|
|
Manufacturer Narrative
|
The recipient's device was explanted.The recipient was reimplanted with another advanced bionics cochlear device.Disclaimer: advanced bionics does not intend that this report be any admission of liability, fault or product defect.
|
|
Manufacturer Narrative
|
Advanced bionics considers the investigation into this reportable event as closed.The external visual inspection revealed that the array was severed prior to receipt, as well as sliced silicone overmold on the top and bottom cover and fantail region.These anomalies are believed to have occurred during revision surgery.The device passed the photographic imaging inspection.System lock was verified.The device passed the electrical tests performed.Advanced electrical testing performed on the device showed lower than typical range on some of the electrodes.The device passed the mechanical test performed.The failure of this device is attributed to an electrode short in the electrode pocket.A corrective action was implemented.This version of the ultra is no longer distributed.This is the final report.Disclaimer: advanced bionics does not intend that this report be any admission of liability, fault or product defect.
|
|
Manufacturer Narrative
|
The recipient was reportedly reimplanted with another cochlear device.Disclaimer: advanced bionics does not intend that this report be any admission of liability, fault or product defect.
|
|