Catalog Number A1114 |
Device Problem
Loose or Intermittent Connection (1371)
|
Patient Problem
No Known Impact Or Consequence To Patient (2692)
|
Event Type
malfunction
|
Manufacturer Narrative
|
An investigation has been initiated based on the reported information.Upon completion of the investigation, a follow-up report will be submitted.
|
|
Event Description
|
A facility reported the standard infinity skull clamp (a1114) would not tighten and loosens when locked.It is unknown if there was patient involvement or if the device failure led to patient injury or surgical delay.
|
|
Manufacturer Narrative
|
Unique device identifier (udi): (b)(4).Complaint confirmed via inspection of the returned a1114 mayfield skull clamp.Unit was received with the lock having rotational and lateral movement and the unit required replacement of worn internal parts.This unit is beyond integra¿s 7 years recommended life cycle (manufactured in 2008).No further investigation required based on the acceptability of risk and no adverse trends identified.This will be monitored and trended going forward.
|
|
Event Description
|
N/a.
|
|