EDWARDS LIFESCIENCES PR SWAN-GANZ CCOMBO V CCO/SVO2/CEDV THERMODILUTION CATHETE; CATHETER, OXIMETER, FIBEROPTIC Back to Search Results
Model Number 774F75
Device Problem Material Rupture (1546)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 12/01/2020
Event Type  malfunction  
Manufacturer Narrative
One 774f75 catheter with a 1.5 syringe and a non-ew contamination shield were returned for examination.The reported event of balloon deflation issue could not be confirmed as per condition of return, however the customer report of balloon ruptured was confirmed.As received, the balloon was found to be ruptured and the rupture edge did not appear to match up.No other visible damage was observed from the catheter body or the returned syringe.All through lumens were patent without any leakage or occlusion.Lot number was not provided, therefore review of the manufacturing records could not be completed.An investigation has been initiated to consider any potential manufacturing factors that may have contributed to this complaint.Invasive procedures involve some patient risks.Although serious complications are relatively uncommon, the physician is advised, before deciding to insert or use the catheter, to consider the potential benefits in relation to the possible complications.The techniques for insertion, methods of using the catheter to obtain patient data information, and the occurrence of complications is well described in the literature.It is standard practice to check balloon integrity by inflating it to the recommended volume in order to detect any asymmetry or leakage condition before use of the catheter.When there is separation of the balloon or fragments from the pulmonary artery catheter, the retained fragment will embolize to the lungs.Due to the large surface area of the pulmonary vasculature, this is generally well tolerated, but can lead to complications such as infection or small infarction.Pulmonary complications may result from improper inflation technique.To avoid damage to the pulmonary artery and possible balloon rupture, the balloon should not be inflated above the recommended volume.Complaint histories for all reported events are reviewed against trending control limits on a monthly basis, and any excursions above the control limits are assessed and documented as part of this monthly review.
 
Event Description
It was reported that during use in a patient of this swan-ganz catheter, the balloon could not be deflated with the syringe attached to the gatevalve.The customer pulled it back and the balloon ruptured.It was specified that this swan ganz was tested before use with patient and there was no issue.There was no vessel injury due to the incident.There was no allegation of patient injury.The device was available for evaluation.Patient demographics were unable to be obtained.
 
Manufacturer Narrative
As part of the manufacturing process, a 100% of the units go through a balloon inflation and visual inspection process.Personnel acknowledgement was conducted at the manufacturing site to prevent recurrence of this type of complaint.A review of the manufacturing records indicated that the product met specifications upon release.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SWAN-GANZ CCOMBO V CCO/SVO2/CEDV THERMODILUTION CATHETE
Type of Device
CATHETER, OXIMETER, FIBEROPTIC
Manufacturer (Section D)
EDWARDS LIFESCIENCES PR
state rd indus pk 402 km 1.4
anasco PR
MDR Report Key11121512
MDR Text Key227070942
Report Number2015691-2021-00020
Device Sequence Number1
Product Code DQE
Combination Product (y/n)N
PMA/PMN Number
K040287
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Type of Report Initial,Followup
Report Date 12/01/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received01/05/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date05/18/2022
Device Model Number774F75
Device Catalogue Number774F75
Device Lot Number63144365
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/21/2020
Date Manufacturer Received02/19/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-