MAUDE Adverse Event Report: ST. JUDE MEDICAL, INC.(CRM-SYLMAR) QUARTET LEAD SMALL-S, 86 CM; PERMANENT PACEMAKER ELECTRODE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ST. JUDE MEDICAL, INC.(CRM-SYLMAR) QUARTET LEAD SMALL-S, 86 CM; PERMANENT PACEMAKER ELECTRODE Back to Search Results
Model Number 1456Q/86
Device Problem Therapy Delivered to Incorrect Body Area (1508)
Patient Problem Undesired Nerve Stimulation (1980)
Event Date 12/17/2020
Event Type  malfunction  
Manufacturer Narrative
This product is registered as a combination product.The results of the investigation are inconclusive since the device was not returned for analysis.Based on the information received, the cause of the reported incident could not be conclusively determined.
 
Event Description
It was reported that the patient experienced phrenic nerve stimulation (pns) and presented in clinic.Upon interrogation, it appeared that there was pns on all vectors of the left ventricular lead.The lead was deactivated.During a second in-clinic visit on (b)(6) 2020, a pacing vector without pns was found and the lead was reprogrammed.The patient was in stable condition.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
QUARTET LEAD SMALL-S, 86 CM
Type of Device
PERMANENT PACEMAKER ELECTRODE
Manufacturer (Section D)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer (Section G)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer Contact
sonali arangil
15900 valley view court
sylmar, CA 91342
MDR Report Key11121781
MDR Text Key225252508
Report Number2017865-2021-00746
Device Sequence Number1
Product Code NIK
UDI-Device Identifier05414734510158
UDI-Public05414734510158
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P030054
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Physician
Type of Report Initial
Report Date 01/05/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date06/30/2023
Device Model Number1456Q/86
Device Catalogue Number1456Q/86
Device Lot NumberA000099050
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 12/17/2020
Initial Date FDA Received01/06/2021
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured07/17/2020
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Age72 YR
Patient Weight113
-
-