Reportable based on device analysis completed on 22dec2020.It was reported that the device could not be inflated.The 70% stenosed target lesion area was located in a mildly tortuous and severely calcified middle section of the right coronary artery.A 10/2.75 flextome cutting balloon was selected for use.During the procedure, it was noted that the device failed to cross the lesion and did not inflate.The device was simply pulled out and the procedure was completed with another of the same device.There were no complications reported and the patient is stable.However, device analysis revealed a balloon pinhole located approximately 1mm proximal of the proximal markerband.
|