• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SMITH & NEPHEW, INC. VULCAN GENERATOR CE MARK; ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITH & NEPHEW, INC. VULCAN GENERATOR CE MARK; ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES Back to Search Results
Model Number 7210812
Device Problem Failure to Power Up (1476)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 12/14/2020
Event Type  malfunction  
Manufacturer Narrative
(b)(4).
 
Event Description
It was reported that the vulcan generator was not turning on.This was noticed during set up the procedure; therefore, no patient was involved.A backup device was available and no delay was reported.All available information has been disclosed.If additional information should become available, a supplemental report will be submitted accordingly.
 
Manufacturer Narrative
H3, h6; the reported device, intended for use in treatment, was received for evaluation.A visual inspection was performed on the product and no issue was observed.There was a relationship found between the returned device and the reported incident.Product failed functional testing with power up failure.Cause of power failure is an unplugged main power harness from base board to the power supply.Product passed functional testing after power harness was reseated to power supply connector.The complaint was confirmed and the root cause has been determined to be an unplugged power harness.A review of the device history records showed there were no indications to suggest that the product did not meet manufacturing specification or would not be able to perform as intended.A complaint history review concluded this was a repeat issue.No containment or corrective actions are recommended at this time.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
VULCAN GENERATOR CE MARK
Type of Device
ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES
Manufacturer (Section D)
SMITH & NEPHEW, INC.
76 s. meridian ave.
oklahoma city OK 73107
MDR Report Key11131390
MDR Text Key225522348
Report Number1643264-2021-00061
Device Sequence Number1
Product Code GEI
UDI-Device Identifier03596010519399
UDI-Public03596010519399
Combination Product (y/n)N
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative
Type of Report Initial,Followup
Report Date 01/27/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number7210812
Device Catalogue Number7210812
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/28/2020
Initial Date Manufacturer Received 12/16/2020
Initial Date FDA Received01/07/2021
Supplement Dates Manufacturer Received01/22/2021
Supplement Dates FDA Received01/27/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-