MAUDE Adverse Event Report: BECTON, DICKINSON & CO. (SPARKS) BD BBL CHOCOLATE II AGAR (GC II AGAR WITH HEMOGLOBIN AND ISOVITALEX); CULTURE MEDIA, FOR ISOLATION OF PATHOGENIC NEISSERIA

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BECTON, DICKINSON & CO. (SPARKS) BD BBL CHOCOLATE II AGAR (GC II AGAR WITH HEMOGLOBIN AND ISOVITALEX); CULTURE MEDIA, FOR ISOLATION OF PATHOGENIC NEISSERIA Back to Search Results
Model Number 221267
Device Problem Incorrect, Inadequate or Imprecise Result or Readings (1535)
Patient Problems No Consequences Or Impact To Patient (2199); No Clinical Signs, Symptoms or Conditions (4582)
Event Date 01/03/2021
Event Type  malfunction  
Manufacturer Narrative
A device evaluation is anticipated but has not yet begun.Upon completion of the investigation, a supplemental report will be filed.(b)(4).
 
Event Description
It was reported that while using the bd bbl¿ chocolate ii agar (gc ii agar with hemoglobin and isovitalex¿) contamination was observed prior to use.The following information was provided by the initial reporter: it was reported that the customer observed microbial growth outside of the streaking area after incubation of inoculated chocolate ii agar plates from lot 0267381 (item 221267).
 
Manufacturer Narrative
H.6.Investigation: during manufacturing of material 221267, media is formulated and sent through a high temperature short time sterilizer to remove bioburden.The petri dishes are subjected to uv radiation to decrease bioburden.The petri dishes are filled in a positive pressure hepa filtered environment.The filled plates are cooled and immediately wrapped into sleeves to decrease the introduction of microbes.Sleeves are then packaged into cartons and then transferred to a refrigerated truck (2 to 8 degrees c) for shipment to the distributor.Bd distributors are provided with the storage guidelines for the shipping and handling of bd media of 2 to 8 degrees c in a dark place.The batch history record for batch 0267381 was satisfactory and no quality notifications were generated during manufacturing and inspection.The release testing that is performed on this product does include bioburden testing.A sample of plates are incubated at 25 degrees c and at 35 degrees c for approximately 72 hours.All bioburden testing performed on this batch was satisfactory per bd internal procedures.Affected product does not have any sterility claims; the product is.Tested for bioburden prior to release to ensure that it conforms to product specifications.However, this does not ensure that the end-user will not receive a contaminated plate.The complaint history was reviewed, and no other complaints have been taken on batch 0267381.Retention samples from batch 0267381 were not available for inspection.No returns or photos were received for investigation.This complaint cannot be confirmed.Bd will continue to trend complaints for contamination.
 
Event Description
It was reported that while using the bd bbl¿ chocolate ii agar (gc ii agar with hemoglobin and isovitalex¿) contamination was observed prior to use the following information was provided by the initial reporter: it was reported that the customer observed microbial growth outside of the streaking area after incubation of inoculated chocolate ii agar plates from lot 0267381 (item 221267).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BD BBL CHOCOLATE II AGAR (GC II AGAR WITH HEMOGLOBIN AND ISOVITALEX)
Type of Device
CULTURE MEDIA, FOR ISOLATION OF PATHOGENIC NEISSERIA
Manufacturer (Section D)
BECTON, DICKINSON & CO. (SPARKS)
7 loveton circle
sparks MD 21152
MDR Report Key11135341
MDR Text Key228214324
Report Number1119779-2021-00040
Device Sequence Number1
Product Code JTY
UDI-Device Identifier10382902212673
UDI-Public10382902212673
Combination Product (y/n)N
PMA/PMN Number
PREAMENDMENT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,other,user facility
Type of Report Initial,Followup
Report Date 08/27/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received01/07/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Expiration Date01/19/2021
Device Model Number221267
Device Catalogue Number221267
Device Lot Number0267381
Was Device Available for Evaluation? No
Date Manufacturer Received08/27/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-