This report is being supplemented to provide additional information based on the legal manufacturer's final investigation.The following sections were updated: g4, g7, h2, h4, h6 and h10.The customer replaced the power switch.Based on the results of the investigation, the event was likely due to the power switch braking down from repeated use for a long period of time, or it was damaged by the user hitting a hard object on it.Handling of the device is described in the instruction manual and this may prevent event, "do not apply excessive force to this video system center and/or other instruments connected.Otherwise, damage and/or malfunction can occur.".
|