Catalog Number 2N3350 |
Device Problem
Fluid/Blood Leak (1250)
|
Patient Problem
No Consequences Or Impact To Patient (2199)
|
Event Date 12/01/2020 |
Event Type
malfunction
|
Manufacturer Narrative
|
Should additional relevant information become available, a supplemental report will be submitted.
|
|
Event Description
|
It was reported, a non-dehp micro-volume extension set leaked.The nurse observed a leak at the filter.The event occurred during an unknown process step.There was no report of patient injury or medical intervention associated with this event.No additional information is available.
|
|
Manufacturer Narrative
|
Additional information: b5, f10 and h6.B5: correction: it was reported, the tube was leaking from the filter; f10: component code added.H10: the device was not returned.And the lot number is unknown.Therefore, a device analysis could not be completed.Should additional relevant information become available, a supplemental report will be submitted.
|
|
Manufacturer Narrative
|
Additional information for h10: based on further investigation, the most probable cause was determined to be related to the end user/methods exceeding the product pressure specifications of 45psi.Should additional relevant information become available, a supplemental report will be submitted.
|
|