TERUMO CARDIOVASCULAR SYSTEMS CORPORATION ADVANCED PERFUSION SYSTEM 1; CONSOLE, HEART-LUNG MACHINE, CARDIOPULMONARY BYPASS- ADVANCED PERFUSION SYSTEM 1 Back to Search Results
Model Number 816300
Device Problem Output Problem (3005)
Patient Problems No Known Impact Or Consequence To Patient (2692); No Clinical Signs, Symptoms or Conditions (4582)
Event Date 12/17/2020
Event Type  malfunction  
Manufacturer Narrative
Per the manufacturer's technical support specialist, at first, the system booted up normally.Once it was rebooted and every time thereafter, the error message was displayed.The field service representative (fsr) verified the disk boot failure.He replaced the coin cell battery, which resolved the issue.The hard drive was due to be replaced in six months.The fsr spoke with the facility's biomedical engineer and it was decided to replace the hard drive as a courtesy and a precaution.All release testing was performed.The unit operated to the manufacturer's specifications.The suspect part was returned to the manufacturer for further evaluation.
 
Event Description
It was reported that during set-up of the device for a cardiopulmonary bypass (cpb) procedure, the central control monitor (ccm) displayed a 'disk boot failure.Insert system disk and press enter' message.As a result, an alternate device was employed.No other details regarding the nature of this event were provided.
 
Manufacturer Narrative
The reported complaint was confirmed.The product surveillance technician (pst) observed the coin battery to be received with 0.375 volts direct current (vdc).The battery did not have enough voltage to supply power to the basic input output system.If additional information becomes available on this complaint that would alter the facts and/or conclusion, a supplemental report will be filed accordingly.
 
Event Description
Additional information received: the surgical procedure was completed successfully.There was no delay, no blood loss, nor adverse consequences to the patient.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ADVANCED PERFUSION SYSTEM 1
Type of Device
CONSOLE, HEART-LUNG MACHINE, CARDIOPULMONARY BYPASS- ADVANCED PERFUSION SYSTEM 1
Manufacturer (Section D)
TERUMO CARDIOVASCULAR SYSTEMS CORPORATION
6200 jackson road
ann arbor MI 48103
MDR Report Key11148682
MDR Text Key226091079
Report Number1828100-2021-00015
Device Sequence Number1
Product Code DTQ
Combination Product (y/n)N
PMA/PMN Number
K172220
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Type of Report Initial,Followup
Report Date 02/22/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number816300
Device Catalogue Number816300
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/22/2020
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 12/17/2020
Initial Date FDA Received01/11/2021
Supplement Dates Manufacturer Received02/04/2021
Supplement Dates FDA Received02/22/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-