Model Number 397002-001 |
Device Problems
Electrical /Electronic Property Problem (1198); Mechanical Problem (1384)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 12/22/2020 |
Event Type
malfunction
|
Manufacturer Narrative
|
The companion 2 driver will be evaluated by syncardia.The results will be provided in a follow-up mdr.(b)(4) initial.
|
|
Event Description
|
While performing a routine evaluation, a syncardia technician reported that the companion 2 driver vacuum blowers did not function.
|
|
Manufacturer Narrative
|
Corrected section b5.The reported issue of the companion 2 driver not passing the system check was confirmed and reproduced during investigation testing.The root cause was determined to be malfunction of the pressure sensor board.This issue will be monitored and trended as part of the customer complaint process.Syncardia has completed its investigation and is closing this file.(b)(4) follow-up report 1.
|
|
Event Description
|
While performing a routine evaluation, a syncardia technician reported that the companion 2 driver did not pass the system check.
|
|