• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SMITH & NEPHEW, INC. COUNTERSINK FOR 2.0MM W/AO QC; PLATE, FIXATION, BONE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITH & NEPHEW, INC. COUNTERSINK FOR 2.0MM W/AO QC; PLATE, FIXATION, BONE Back to Search Results
Model Number 71174923
Device Problem Dull, Blunt (2407)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 12/18/2020
Event Type  Injury  
Manufacturer Narrative
(b)(4).
 
Event Description
It was reported that, during surgery,screw countersink device was dull and didn´t cut.The device failed during use while inside the patient.A change in surgical technique was made to finished the procedure with a 2.4 countersink and with no delay in the surgery.No other complications were reported.All available information has been disclosed.If additional information should become available, a supplemental report will be submitted accordingly.
 
Manufacturer Narrative
H3, h6: the device, used in treatment, was not returned for evaluation and the reported event could not be confirmed.A review of the manufacturing records did not reveal a manufacturing abnormality that could have caused or contributed to the reported incident.A review of complaint history did not reveal additional complaints for the listed batch.A review of the risk management file revealed this failure mode was previously identified.At this time, we have no reason to suspect that the product failed to meet any product specifications at the time of manufacture.This device is a reusable instrument that can be exposed to numerous surgeries.Damage from prolonged use, misuse or rough handling are likely probable causes of the reported event.We recommend that all reusable instruments be routinely inspected for wear and damage and replaced as necessary.Based on this investigation, the need for corrective action is not indicated.Without the return of the actual product involved, our investigation could not proceed.Should the device or additional information be received, the complaint will be reopened.No further investigation is warranted for this complaint; however, we will continue to monitor for future complaints and investigate as necessary.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
COUNTERSINK FOR 2.0MM W/AO QC
Type of Device
PLATE, FIXATION, BONE
Manufacturer (Section D)
SMITH & NEPHEW, INC.
1450 brooks rd.
memphis TN 38116
MDR Report Key11150876
MDR Text Key226217258
Report Number1020279-2021-00247
Device Sequence Number1
Product Code HRS
UDI-Device Identifier00885556387443
UDI-Public00885556387443
Combination Product (y/n)N
PMA/PMN Number
K132886
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Type of Report Initial,Followup
Report Date 09/03/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number71174923
Device Catalogue Number71174923
Device Lot Number15BNX0016
Was Device Available for Evaluation? No
Initial Date Manufacturer Received 12/18/2020
Initial Date FDA Received01/11/2021
Supplement Dates Manufacturer Received09/02/2021
Supplement Dates FDA Received09/03/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
-
-