It was reported that during a replacement procedure due to normal battery depletion, the physician experienced difficulty in tightening the setscrew of this device to the right ventricular (rv) lead.The physician elected to exchange this device to resolve the event and the procedure was successfully completed.The patient was stable with no adverse consequences.The device is expected for analysis, but has not yet been received.
|
The returned device was thoroughly inspected and analyzed upon receipt at our post market quality assurance laboratory.Visual examination of the device header and case noted a missing rv setscrew.Pin gauge testing, designed to verify proper port dimensions, was completed.Each port measured as expected.The device was then exposed to simulated heart load conditions, and the defibrillation, pacing, and sensing functions were tested.Impedance testing was completed and all measurements were within normal limits.The device operated appropriately with no interruptions in therapy output at the returned programmed settings.A series of electrical tests was also performed, and again, normal device function was observed.Analysis did not identify any device characteristics that would have caused or contributed to the missing rv setscrew or reported tightening difficulties.
|