Model Number 72404013 |
Device Problems
Mechanical Problem (1384); Material Puncture/Hole (1504)
|
Patient Problems
No Consequences Or Impact To Patient (2199); No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 12/22/2020 |
Event Type
Injury
|
Event Description
|
It was reported that the patient underwent a surgical procedure to revise this inflatable penile prosthesis (ipp) due to the device not working properly that started two weeks prior to the procedure.The ipp cylinder and pump were explanted and replaced with a new ipp cylinder and pump.During the revision procedure the physician found that one of the cylinders contained a hole, the hole origination is unknown.It was found that the cylinder did not expand due to saline leakage due to the cylinder hole.The patient was stable following the procedure.
|
|
Manufacturer Narrative
|
The ams700 ipp cylinder was visually inspected and functionally tested.Cylinder has a leak in the distal cylinder body that was the result of sharp instrument damage consistent with explant damage; hole was present.Due to this damage being consistent with explant it will be considered a secondary failure.Cylinder has fold that indicate possible buckling of the cylinders in the proximal cylinder body.Cylinder was not pressure test due to leak was identified.The ams 700 momentary squeeze (ms) pump was visually inspected; no leaks were found.The tubing was identified to be cut down to the pump block.The tubing was not returned for analysis.The pump was functionally tested and failed the 8lb.Activation test.The pump therefore required more than 8lbs.Of force to activate.Product analysis was unable to confirm the reported event.
|
|
Event Description
|
It was reported that the patient underwent a surgical procedure to revise this inflatable penile prosthesis (ipp) due to the device not working properly that started two weeks prior to the procedure.The ipp cylinder and pump were explanted and replaced with a new ipp cylinder and pump.During the revision procedure the physician found that one of the cylinders contained a hole, the hole origination is unknown.It was found that the cylinder did not expand due to saline leakage due to the cylinder hole.The patient was stable following the procedure.
|
|
Search Alerts/Recalls
|