The unit was inspected and tested.During the start up testing, the unit showed error message¿ remove instrument¿ due to faulty pkrf and auxiliary boards.Inspection of the unit found the id and pkrf boards were not properly mounted, the 3 stands on pkrf board were observed to be broken.The output testing was found low due to to faulty psu (power supply unit).The unit is running an old version software v2.04 and needs to be upgraded to v2.07.Review of the fault log shows 400 ref 25, seven times indicated that the biomed time credit expired.The biomed test cable has been used beyond its time limit allocation.Error 100 ref 12, showed one time indicated a non volatile memory corrupt or not initialized.Investigation is ongoing therefore the root cause cannot be determined at this time.If additional information becomes available this report will be supplemented accordingly.
|