• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ST. JUDE MEDICAL, INC.(CRM-SYLMAR) GALLANT DR; IMPLANTABLE CARDIOVERTER DEFIBRILLATOR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ST. JUDE MEDICAL, INC.(CRM-SYLMAR) GALLANT DR; IMPLANTABLE CARDIOVERTER DEFIBRILLATOR Back to Search Results
Model Number CDDRA500Q
Device Problem Failure to Interrogate (1332)
Patient Problems No Consequences Or Impact To Patient (2199); No Clinical Signs, Symptoms or Conditions (4582)
Event Date 12/23/2020
Event Type  malfunction  
Manufacturer Narrative
The results/method and conclusion codes along with investigation results will be provided in the final report.
 
Event Description
It was reported that the implantable cardioverter defibrillator (icd) was being prepared for implant.During preparation, it was found that the icd was failing to be connected and interrogated via bluetooth prior to implant.The device was withheld from installation and an alternate device was used.There was no patient involvement.
 
Manufacturer Narrative
Analysis was normal.No anomalies were found.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
GALLANT DR
Type of Device
IMPLANTABLE CARDIOVERTER DEFIBRILLATOR
Manufacturer (Section D)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
MDR Report Key11155997
MDR Text Key226312061
Report Number2017865-2021-01994
Device Sequence Number1
Product Code NIK
UDI-Device Identifier05415067032003
UDI-Public05415067032003
Combination Product (y/n)N
PMA/PMN Number
P030054
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Type of Report Initial,Followup
Report Date 02/24/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date12/31/2020
Device Model NumberCDDRA500Q
Device Catalogue NumberCDDRA500Q
Device Lot NumberS000077185
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/04/2021
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 12/23/2020
Initial Date FDA Received01/12/2021
Supplement Dates Manufacturer Received02/19/2021
Supplement Dates FDA Received02/24/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-