• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ABBOTT MEDICAL KIT IMPLANTABLE SLIM TIP LEAD, 50CM; DRG LEAD

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ABBOTT MEDICAL KIT IMPLANTABLE SLIM TIP LEAD, 50CM; DRG LEAD Back to Search Results
Model Number MN10450-50A
Device Problem Fracture (1260)
Patient Problems No Consequences Or Impact To Patient (2199); No Clinical Signs, Symptoms or Conditions (4582)
Event Date 12/15/2020
Event Type  Injury  
Manufacturer Narrative
The results/method and conclusion codes along with investigation results will be provided in the final report.Date of event is estimated.
 
Event Description
Related manufacturer reference number: 1627487-2021-00044.It was reported that, while adding 2 leads, the physician discovered that two of the patient's implanted leads were fractured.As a result, the patient's leads were explanted and replaced.
 
Manufacturer Narrative
The report event of fracture lead was confirmed.As received, visual inspection showed the returned lead (b) found a damaged on the outer tubing and all the internal lead wires being broken.These fractures are consistent with an overstress condition or sudden event the lead was subjected to while still in vivo.The cause of the event is consistent with damage during use.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
KIT IMPLANTABLE SLIM TIP LEAD, 50CM
Type of Device
DRG LEAD
Manufacturer (Section D)
ABBOTT MEDICAL
6901 preston rd
plano TX 75024
MDR Report Key11156352
MDR Text Key226312784
Report Number1627487-2021-00043
Device Sequence Number1
Product Code PMP
UDI-Device Identifier05415067027153
UDI-Public05415067027153
Combination Product (y/n)N
PMA/PMN Number
P150004
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Type of Report Initial,Followup
Report Date 02/05/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date08/15/2021
Device Model NumberMN10450-50A
Device Catalogue NumberMN10450-50A
Device Lot Number7113424
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/20/2021
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 12/15/2020
Initial Date FDA Received01/12/2021
Supplement Dates Manufacturer Received02/05/2021
Supplement Dates FDA Received02/05/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Treatment
MN10450-50A; MN10450-50A
Patient Outcome(s) Other;
Patient Age34 YR
Patient Weight48
-
-