• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: NUVASIVE, INCORPORATED NUVASIVE RELINE SYSTEM; THORACOLUMBOSACRAL PEDICLE SCREW SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

NUVASIVE, INCORPORATED NUVASIVE RELINE SYSTEM; THORACOLUMBOSACRAL PEDICLE SCREW SYSTEM Back to Search Results
Model Number 13116540
Device Problem Loosening of Implant Not Related to Bone-Ingrowth (4002)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 12/21/2020
Event Type  malfunction  
Manufacturer Narrative
No product was returned for evaluation due to product being discarded at the facility.No images were provided to confirm the product malfunction.It is unknown if the patient followed post-operative restrictions or suffered a fall.Review of reported event identified nuvasive posterior fixation was utilized with a migrated competitor's interbody and most likely the root cause.Combining the use of nuvasive products and other manufacturer's is not recommended as they are untested and results may vary.No additional investigation can be completed.Labeling review: ".Potential adverse events and complications: as with any major surgical procedures, there are risks involved in orthopedic surgery.Potential risks identified with the use of this system, which may require additional surgery, include: bending, fracture or loosening of implant component(s), loss of fixation, nonunion or delayed union.".".Warnings, cautions and precautions: these devices can break when subjected to the increased load associated with delayed union or nonunion.Internal fixation appliances are load-sharing devices that hold bony structures in alignment until healing occurs.If healing is delayed, or does not occur, the implant may eventually loosen, bend, or break.Loads on the device produced by load bearing and by the patient¿s activity level will dictate the longevity of the implant.".".Patient education: the patient should be made aware of the limitations of the implant and potential risks of the surgery.The patient should be instructed to limit postoperative activity, as this will reduce the risk of bent, broken or loose implant components.The patient must be made aware that implant components may bend, break or loosen even though restrictions in activity are followed.".".Compatibility: do not use the reline system with components of other systems, other than the armada system.Refer to the armada system instructions for use for a list of the armada system indications for use.Unless stated otherwise, nuvasive devices are not to be combined with the components of another system.".".Post-operative warnings: damage to the weight-bearing structures can give rise to loosening of the components, dislocation and migration, as well as to other complications.".
 
Event Description
On (b)(6) 2019 a patient underwent a posterior lumbar interbody fusion with a competitor's cage and nuvasive products at l4/5.On (b)(6) 2020 a revision surgery was performed due to the competitor's cage backing out.It was also found that the nuvasive posterior fixation had loosened, therefore the fixation and competitor's cage was replaced.Non-union was observed as well.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
NUVASIVE RELINE SYSTEM
Type of Device
THORACOLUMBOSACRAL PEDICLE SCREW SYSTEM
Manufacturer (Section D)
NUVASIVE, INCORPORATED
7475 lusk boulevard
san diego CA 92121
Manufacturer (Section G)
NUVASIVE, INCORPORATED
7475 lusk boulevard
san diego CA 92121
Manufacturer Contact
virnalisa cobacha
7475 lusk boulevard
san diego, CA 92121
MDR Report Key11158625
MDR Text Key227784210
Report Number2031966-2021-00004
Device Sequence Number1
Product Code NKB
UDI-Device Identifier00887517582492
UDI-Public887517582492
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
K143684
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Physician
Type of Report Initial
Report Date 01/12/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number13116540
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 12/24/2020
Initial Date FDA Received01/13/2021
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-