No product was returned for evaluation due to product being discarded at the facility.No images were provided to confirm the product malfunction.It is unknown if the patient followed post-operative restrictions or suffered a fall.Review of reported event identified nuvasive posterior fixation was utilized with a migrated competitor's interbody and most likely the root cause.Combining the use of nuvasive products and other manufacturer's is not recommended as they are untested and results may vary.No additional investigation can be completed.Labeling review: ".Potential adverse events and complications: as with any major surgical procedures, there are risks involved in orthopedic surgery.Potential risks identified with the use of this system, which may require additional surgery, include: bending, fracture or loosening of implant component(s), loss of fixation, nonunion or delayed union.".".Warnings, cautions and precautions: these devices can break when subjected to the increased load associated with delayed union or nonunion.Internal fixation appliances are load-sharing devices that hold bony structures in alignment until healing occurs.If healing is delayed, or does not occur, the implant may eventually loosen, bend, or break.Loads on the device produced by load bearing and by the patient¿s activity level will dictate the longevity of the implant.".".Patient education: the patient should be made aware of the limitations of the implant and potential risks of the surgery.The patient should be instructed to limit postoperative activity, as this will reduce the risk of bent, broken or loose implant components.The patient must be made aware that implant components may bend, break or loosen even though restrictions in activity are followed.".".Compatibility: do not use the reline system with components of other systems, other than the armada system.Refer to the armada system instructions for use for a list of the armada system indications for use.Unless stated otherwise, nuvasive devices are not to be combined with the components of another system.".".Post-operative warnings: damage to the weight-bearing structures can give rise to loosening of the components, dislocation and migration, as well as to other complications.".
|