Model Number CD3369-40Q |
Device Problem
Over-Sensing (1438)
|
Patient Problem
No Consequences Or Impact To Patient (2199)
|
Event Date 12/29/2020 |
Event Type
malfunction
|
Manufacturer Narrative
|
The results of the investigation are inconclusive since the device was not returned for analysis.Based on the information received, the cause of the reported incident could not be conclusively determined.
|
|
Event Description
|
It was reported that the asymptomatic patient presented remotely.Review of the transmission revealed non-sustained right ventricular over-sensing episodes due to post-paced t-wave over-sensing.Programming changes were discussed; no intervention was reported.The patient was in stable condition.
|
|
Event Description
|
New information received noted that the programming changes were made on 19 jan 2021.
|
|