Additional information was added to d10, h3, h4 and h6.Additional information/correction: b5: (missing information); f10/h6: device codes (add 1250).B5: a leak was observed, as a result of the separation; h4: the lot was manufactured from august 10, 2020 - august 11, 2020.H10: the device was received for evaluation.Unaided visual inspection was performed, which observed that the spike was detached.No adhesive could be observed, at the spike tubing end upon magnified inspection.Due to the nature of the returned sample, no additional testing could be performed.The reported condition was verified.The cause of the detached spike could not be determined.However the most likely cause, was due to inadequate or lack of adhesive applied, during manufacturing.A batch review was conducted, and there were no deviations found related to this reported condition, during the manufacture of this lot.Should additional relevant information become available, a supplemental report will be submitted.
|