Catalog Number H93811
Device Problems Disconnection (1171); Fluid/Blood Leak (1250)
Patient Problem No Patient Involvement (2645)
Event Date 12/11/2020
Event Type  malfunction  
Manufacturer Narrative
Should additional relevant information become available, a supplemental report will be submitted.
 
Event Description
It was reported that the spike of a sterile repeater pump tube set separated from the tubing when attaching it to a fluid filled bag.This issue was identified during setup and preparation prior to patient use.There was no patient involvement.No additional information is available.
 
Manufacturer Narrative
Additional information was added to d10, h3, h4 and h6.Additional information/correction: b5: (missing information); f10/h6: device codes (add 1250).B5: a leak was observed, as a result of the separation; h4: the lot was manufactured from august 10, 2020 - august 11, 2020.H10: the device was received for evaluation.Unaided visual inspection was performed, which observed that the spike was detached.No adhesive could be observed, at the spike tubing end upon magnified inspection.Due to the nature of the returned sample, no additional testing could be performed.The reported condition was verified.The cause of the detached spike could not be determined.However the most likely cause, was due to inadequate or lack of adhesive applied, during manufacturing.A batch review was conducted, and there were no deviations found related to this reported condition, during the manufacture of this lot.Should additional relevant information become available, a supplemental report will be submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TUBE SET, STD VOL, STER
Type of Device
SYSTEM/DEVICE, PHARMACY COMPOUNDING
Manufacturer (Section D)
BAXTER HEALTHCARE CORPORATION
deerfield IL
MDR Report Key11162637
MDR Text Key231205684
Report Number1416980-2020-08472
Device Sequence Number1
Product Code NEP
UDI-Device Identifier00085412475721
UDI-Public(01)00085412475721
Combination Product (y/n)Y
PMA/PMN Number
K900585
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,consum
Type of Report Initial,Followup
Report Date 02/11/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received01/13/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Expiration Date07/31/2023
Device Catalogue NumberH93811
Device Lot Number60251707
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/05/2021
Date Manufacturer Received02/01/2021
Is This a Reprocessed and Reused Single-Use Device? No