Model Number CD3369-40Q |
Device Problems
Loose or Intermittent Connection (1371); Misconnection (1399)
|
Patient Problems
No Consequences Or Impact To Patient (2199); No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 12/29/2020 |
Event Type
malfunction
|
Manufacturer Narrative
|
The results/method and conclusion codes along with investigation results will be provided in the final report.
|
|
Event Description
|
It was reported that the patient presented for an implant procedure.During the procedure, when the physician removed the left ventricular lead, the set screw would not go back into place as it had retracted too far.The device was removed and replaced.The patient was in stable condition.
|
|
Manufacturer Narrative
|
The reported inability to tighten the left ventricle set screw was confirmed in the laboratory.Visual inspection identified septum material inside the left ventricle screw hex cavity and the set screw was stripped.This material prevented full insertion of the torque driver in the hex cavity and resulted in difficulty tightening the set screw.The cause was likely due to the handling of the device during implant.There was no device malfunction.
|
|