• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION ULTRAFLEX TRACHEOBRONCHIAL; PROSTHESIS, TRACHEAL, EXPANDABLE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION ULTRAFLEX TRACHEOBRONCHIAL; PROSTHESIS, TRACHEAL, EXPANDABLE Back to Search Results
Model Number M00564860
Device Problems Break (1069); Unraveled Material (1664); Positioning Problem (3009)
Patient Problems Injury (2348); Insufficient Information (4580)
Event Date 12/22/2020
Event Type  Injury  
Manufacturer Narrative
(b)(4).The device has been received for analysis.Upon completion of the failure analysis of the complaint device, if there is any further relevant information from that review, a supplemental medwatch will be filed.
 
Event Description
It was reported to boston scientific corporation on december 23, 2020 that an ultraflex tracheobronchial distal release covered stent was to be implanted to treat a malignant airway tumor in the main trachea during a tracheal stent implantation procedure performed on (b)(6) 2020.Reportedly, the patient's anatomy was not tortuous prior to stent placement.According to the complainant, during the procedure, the stent was fully deployed inside the patient but in an incorrect location.The physician pulled the green retention suture to adjust the position of the stent but the stent suture broke and the stent unraveled.The stent was removed from the patient and the procedure was rescheduled.Reportedly, the patient was hospitalized for observation 1 to 2 weeks until another procedure was performed and a non-bsc device was implanted.Reportedly, the patient had minor airway mucosa damage that did not require any medical intervention.The patient's condition at the conclusion of the procedure was reported to be stable.
 
Manufacturer Narrative
Block h6: medical device problem code a0401 captures the reportable event of stent suture break.Medical device problem code a1502 captures the reportable event of stent positioning issue.Medical device problem code a040508 captures the reportable event of stent unraveled material.Block h10: a deployed ultraflex tracheobronchial covered stent and deployment suture were received for analysis; the delivery system was not returned.Visual examination of the returned device found the stent was returned unraveled and the green suture was broken.The stent length was measured and was found not to be within specification as the stent was returned unraveled.No other issues with the stent was noted.The reported event of stent suture break and stent unraveled material were confirmed during visual examination.The investigation concluded that the reported event and the observed failures were likely due to factors encountered during the procedure.It may be that the technique used by the user and the incorrect location of the deployed stent could have contributed to the stent suture to break and the stent to unravel when the physician likely pulled the green retention suture to adjust the position of the stent during the procedure.Therefore, a review and analysis of all available information indicated the most probable cause is adverse event related to procedure.A review of the manufacturing documentation for this device revealed that no anomalies or deviations related to the event occurred during manufacturing.A labeling review was performed, and from the information available, this device was used per the instructions for use (ifu)/ product label.
 
Event Description
It was reported to boston scientific corporation on (b)(6) 2020 that an ultraflex tracheobronchial distal release covered stent was to be implanted to treat a malignant airway tumor in the main trachea during a tracheal stent implantation procedure performed on (b)(6) 2020.Reportedly, the patient's anatomy was not tortuous prior to stent placement.According to the complainant, during the procedure, the stent was fully deployed inside the patient but in an incorrect location.The physician pulled the green retention suture to adjust the position of the stent but the stent suture broke and the stent unraveled.The stent was removed from the patient and the procedure was rescheduled.Reportedly, the patient was hospitalized for observation 1 to 2 weeks until another procedure was performed and a non-bsc device was implanted.Reportedly, the patient had minor airway mucosa damage that did not require any medical intervention.The patient's condition at the conclusion of the procedure was reported to be stable.Additional information received on january 13, 2021: according to the complainant, the patient was hospitalized prior to the stent placement procedure and remained in the hospital for 1 to 2 weeks where routine observation was performed until another procedure was performed to implant a non-bsc device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ULTRAFLEX TRACHEOBRONCHIAL
Type of Device
PROSTHESIS, TRACHEAL, EXPANDABLE
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
300 boston scientific way
marlborough MA 01752
MDR Report Key11166917
MDR Text Key226741986
Report Number3005099803-2021-00099
Device Sequence Number1
Product Code JCT
Combination Product (y/n)N
PMA/PMN Number
K012883
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,distri
Type of Report Initial,Followup
Report Date 02/04/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date05/28/2022
Device Model NumberM00564860
Device Catalogue Number6486
Device Lot Number0025542731
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/12/2021
Initial Date Manufacturer Received 12/23/2020
Initial Date FDA Received01/13/2021
Supplement Dates Manufacturer Received01/13/2021
Supplement Dates FDA Received02/04/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
Patient Age57 YR
Patient Weight84
-
-