• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DENTSPLY VDW GMBH VDW GOLD/SILVER CONTRA ANGLE; HANDPIECE, CONTRA- AND RIGHT-ANGLE ATTACHMENT, DENTAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DENTSPLY VDW GMBH VDW GOLD/SILVER CONTRA ANGLE; HANDPIECE, CONTRA- AND RIGHT-ANGLE ATTACHMENT, DENTAL Back to Search Results
Catalog Number V041079000000
Device Problem Mechanical Problem (1384)
Patient Problem Pain (1994)
Event Type  malfunction  
Manufacturer Narrative
Since incorrect measurement could necessitate medical/surgical intervention to preclude permanent damage to a body structure or permanent impairment of a body function, this malfunction would be likely to cause/contribute to a serious injury should it recur.As such, this event meets the criteria for reportability per 21 cfr part 803.Battery (voltage breaks down under load) defective, replaced.Lip clip cable (insulation broken) defective, replaced.Power supply unit, file clamp and file clamp cable were not included.Micromotor gmr1876559, contra-angle handpieces 60947, 25305 and foot control 86850 are not defective.
 
Event Description
In this event it was reported that a contra angle 6:1 for vdw.Gold/silver causes incorrect measurement results when used with an apex locator.According to customer, "during treatment with the device, the patient is injured again and again, as the lack of an apex stop causes over-instrumentation, which then leads to pain.".
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
VDW GOLD/SILVER CONTRA ANGLE
Type of Device
HANDPIECE, CONTRA- AND RIGHT-ANGLE ATTACHMENT, DENTAL
Manufacturer (Section D)
DENTSPLY VDW GMBH
bayerwaldstrasse 15
munich, 81737
GM  81737
Manufacturer (Section G)
SIRONA DENTAL SYSTEMS GMBH
fabrikstrasse 31
bensheim, hessen D-646 25
GM   D-64625
Manufacturer Contact
karl nittinger
221 w. philadelphia st.
suite 60w
york, PA 17401
7178494424
MDR Report Key11168236
MDR Text Key227110896
Report Number9611053-2020-00426
Device Sequence Number1
Product Code EGS
Combination Product (y/n)N
Reporter Country CodeAU
PMA/PMN Number
NA#S-P#NA
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,health professional
Reporter Occupation Dentist
Type of Report Initial
Report Date 01/13/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received01/13/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue NumberV041079000000
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/14/2020
Is the Reporter a Health Professional? Yes
Date Manufacturer Received12/14/2020
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-