Product analysis:analysis was able to confirm the customers comment that the external pulse generator (epg) does not conduct.The battery chamber door was difficult to open, the releasing mechanism was contaminated and sticky.There was blood residue on the battery drawer.Communication with test system was difficult and incoming test was stopped during test.The main board and heart wire contacts were checked.Battery contacts and chamber were inspected and cleaned also.An additional incoming test was performed and failed, the main board was suspected to be defective.The epg is at end of service, customer advised to scrap.Correction: e medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|