Model Number C40+
Device Problems Circuit Failure (1089); Noise, Audible (3273)
Patient Problem Failure of Implant (1924)
Event Date 01/07/2021
Event Type  Injury  
Manufacturer Narrative
The device has not been explanted.If it should be explanted, it is to be returned to the manufacturer for evaluation.When available, a device failure analysis will be submitted as a follow up report.
 
Event Description
The hearing performance using the device is affected.On (b)(6) 2021 the user experienced a short circuit like, inconstant noise, beside the speech signal and therefore stopped to wear the audio processor.The user also reported a previous noise sensation similar to tinnitus in (b)(6) 2020 which was reduced after a fitting adjustment.
 
Event Description
The hearing performance using the device is affected.On (b)(6) 2021 the user experienced a short circuit like, inconstant noise, beside the speechsignal and therefore stopped to wear the audio processor.The user also reported a previous noise sensation similar to tinnitus in (b)(6) 2020 which was reduced after a fitting adjustment.The recipient was re-implanted on (b)(6) 2021.
 
Manufacturer Narrative
Conclusion: device investigation confirmed that the stimulator electronics failed.The failure of the electronic circuitry was probably caused by humidity ingress at detected micro-leaks at the housing braze joint.Other damages found during investigation are most likely related to explantation surgery.The problems described in the recipient report appear to match the damage found.This is a final report.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MED-EL COMBI 40+ COCHLEAR IMPLANT SYSTEM
Type of Device
COCHLEAR IMPLANT
Manufacturer (Section D)
MED-EL ELEKTROMEDIZINISCHE GERAETE GMBH
innsbruck
AU 
MDR Report Key11176271
MDR Text Key228220838
Report Number9710014-2021-00022
Device Sequence Number1
Product Code MCM
Combination Product (y/n)N
PMA/PMN Number
P000025
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative
Type of Report Initial,Followup
Report Date 06/02/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Model NumberC40+
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/05/2021
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 01/13/2021
Initial Date FDA Received01/15/2021
Supplement Dates Manufacturer Received01/13/2021
Supplement Dates FDA Received06/02/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
Patient Age21 YR
-
-