Model Number C40+ |
Device Problems
Circuit Failure (1089); Noise, Audible (3273)
|
Patient Problem
Failure of Implant (1924)
|
Event Date 01/07/2021 |
Event Type
Injury
|
Manufacturer Narrative
|
The device has not been explanted.If it should be explanted, it is to be returned to the manufacturer for evaluation.When available, a device failure analysis will be submitted as a follow up report.
|
|
Event Description
|
The hearing performance using the device is affected.On (b)(6) 2021 the user experienced a short circuit like, inconstant noise, beside the speech signal and therefore stopped to wear the audio processor.The user also reported a previous noise sensation similar to tinnitus in (b)(6) 2020 which was reduced after a fitting adjustment.
|
|
Event Description
|
The hearing performance using the device is affected.On (b)(6) 2021 the user experienced a short circuit like, inconstant noise, beside the speechsignal and therefore stopped to wear the audio processor.The user also reported a previous noise sensation similar to tinnitus in (b)(6) 2020 which was reduced after a fitting adjustment.The recipient was re-implanted on (b)(6) 2021.
|
|
Manufacturer Narrative
|
Conclusion: device investigation confirmed that the stimulator electronics failed.The failure of the electronic circuitry was probably caused by humidity ingress at detected micro-leaks at the housing braze joint.Other damages found during investigation are most likely related to explantation surgery.The problems described in the recipient report appear to match the damage found.This is a final report.
|
|