Model Number CLH-2 |
Device Problems
Electrical /Electronic Property Problem (1198); Power Problem (3010)
|
Patient Problem
No Consequences Or Impact To Patient (2199)
|
Event Type
malfunction
|
Manufacturer Narrative
|
The subject device was not returned to omsc for evaluation.The exact cause of the reported event could not be conclusively determined at this time.If additional information becomes available, this report will be supplemented.
|
|
Event Description
|
Olympus medical systems corp.(omsc) was informed from the user facility that in unspecified timing, it was found that the subject device could not be turned the power on.There was no report of patient injury associated with this event.Olympus service operation repair center (sorc) checked the subject device and found that the reported phenomenon was not duplicated.
|
|
Manufacturer Narrative
|
This supplemental report is being submitted to provide additional information.The subject device was not returned to olympus medical systems corp.(omsc) for evaluation.Omsc reviewed the manufacturing history (dhr) of the subject device and confirmed no irregularity.The exact cause of the reported phenomenon could not be conclusively determined.However, based upon the information from olympus service operation repair center (sorc), there was the possibility that the reported phenomenon was attributed to the following.The power cord was not firmly connected.The power cord was about to break.The lamp access cover was not closed tightly.The power switch of the mobile workstation was not turned on.
|
|