• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ABBOTT MEDICAL SWIFT-LOCK ANCHOR; SCS ANCHOR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ABBOTT MEDICAL SWIFT-LOCK ANCHOR; SCS ANCHOR Back to Search Results
Model Number 1192
Device Problem Fracture (1260)
Patient Problem Inadequate Pain Relief (2388)
Event Date 09/25/2020
Event Type  Injury  
Manufacturer Narrative
During processing of this complaint, attempts were made to obtain complete patient information.The results of the investigation are inconclusive since the device was not returned for analysis.Based on the information received, the cause of the reported incident could not be conclusively determined.
 
Event Description
Related manufacturer reference number: 3006705815-2021-00118.Related manufacturer reference number: 3006705815-2021-00119.Related manufacturer reference number: 1627487-2021-00302.It was reported the patient was experiencing ineffective stimulation and stimulation was decreasing by itself.Diagnostic testing indicated high impedances.Reprogramming was initially able to address the issue; however, patient¿s both leads were found to be migrated.In turn, surgical intervention was undertaken wherein it was discovered that one of the leads was fractured at the anchor site and the anchor was also broken.The lead and anchor were explanted and replaced.The physician decided to place the new lead up higher at t4.The existing lead was also revised to t4.This addressed the issue.It is unknown which anchor was broken and was explanted, therefore both anchors are being reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SWIFT-LOCK ANCHOR
Type of Device
SCS ANCHOR
Manufacturer (Section D)
ABBOTT MEDICAL
6901 preston rd
plano TX 75024
Manufacturer (Section G)
ABBOTT MEDICAL
6901 preston rd
plano TX 75024
Manufacturer Contact
ronnie shalev
6901 preston road
plano, TX 75024
9723098000
MDR Report Key11177062
MDR Text Key227011837
Report Number1627487-2021-00303
Device Sequence Number1
Product Code GZB
UDI-Device Identifier05415067024084
UDI-Public05415067024084
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K092371
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,consum
Reporter Occupation Physician
Type of Report Initial
Report Date 01/15/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received01/15/2021
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date02/07/2021
Device Model Number1192
Device Catalogue Number1192
Device Lot Number6840025
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received12/30/2020
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured02/08/2019
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
1192, SCS ANCHOR; 3186 X 2, SCS LEAD
Patient Outcome(s) Other;
-
-