BOSTON SCIENTIFIC CORPORATION AUTOGEN; DEFIBRILLATOR, AUTOMATIC IMPLANTABLE CARDIOVERTER, WITH CRT-D Back to Search Results
Model Number G173
Device Problems High impedance (1291); Off-Label Use (1494)
Patient Problem No Code Available (3191)
Event Date 10/01/2020
Event Type  malfunction  
Manufacturer Narrative
(b)(4) is being used to capture the prolonged procedure required to replace this lead during initial implant.This device has not been returned; therefore, technical analysis cannot be conducted.Without a returned device it is not possible to definitively confirm how the device may have contributed to the complaint incident.If information is provided in the future, a supplemental report will be issued.
 
Event Description
It was reported that during implant this cardiac resynchronization therapy defibrillator (crt-d) exhibited high out of range shock impedance of greater than 200 ohms during shock configuration.The shock impedance was fine with the previous device.A different crt-d device was replaced and the shock configuration was successful.Another synchronous shock test was performed on the first crt-d and it was now possible to perform shock but still exhibited high shock impedance while in pocket.The physician elected to implant the first crt-d instead with therapy turned off.After further testing, the impedance showed good measurements.This device remains in service.The physician will continue to monitor.No adverse patient effects were reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AUTOGEN
Type of Device
DEFIBRILLATOR, AUTOMATIC IMPLANTABLE CARDIOVERTER, WITH CRT-D
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
4100 hamline avenue north
saint paul MN 55112
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
cashel road
clonmel
EI  
Manufacturer Contact
timothy degroot
4100 hamline avenue north
saint paul, MN 55112
6515826168
MDR Report Key11181986
MDR Text Key227515722
Report Number2124215-2020-28152
Device Sequence Number1
Product Code NIK
Combination Product (y/n)N
Reporter Country CodeGM
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Physician
Type of Report Initial
Report Date 01/15/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received01/15/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Expiration Date09/03/2022
Device Model NumberG173
Device Catalogue NumberG173
Device Lot Number142141
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received10/01/2020
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured09/16/2020
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-