OLYMPUS MEDICAL SYSTEMS CORP. ULTRASONIC PROBE Back to Search Results
Model Number UM-G20-29R-3
Device Problem Fluid/Blood Leak (1250)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Type  malfunction  
Manufacturer Narrative
The subject device was returned to omsc for evaluation.Omsc reviewed the manufacturing history (dhr) of the subject device and confirmed no irregularity.Omsc checked the subject device and found that the ultrasonic propagation fluid medium leaked from the distal end and that the insertion tube was torn.Based on the result of the omsc evaluation, omsc concluded that the reported phenomenon was caused by a tear of the insertion tube and the leakage of the ultrasonic propagation fluid medium leaking through the tear.The exact cause of the tear of the insertion tube could not be conclusively determined, but there was the possibility that it was attributed to the external force such as the hit against the instrument channel port of the endoscope, and so on.If additional information becomes available, this report will be supplemented.
 
Event Description
Olympus medical systems corp.(omsc) was informed from the user facility that during the preparation for an unspecified procedure using the subject device, it was found that the ultrasonic propagation fluid medium leaked from the distal end of the subject device.The subject device had been used twice, and the reported event occurred during the third preparation for use.There was no report of patient injury associated with this event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ULTRASONIC PROBE
Type of Device
ULTRASONIC PROBE
Manufacturer (Section D)
OLYMPUS MEDICAL SYSTEMS CORP.
2951 ishikawa-cho
hachioji-shi, tokyo-to 192-8 507
JA  192-8507
Manufacturer Contact
kazutaka matsumoto
2951 ishikawa-cho
hachioji-shi, tokyo-to 192-8-507
JA   192-8507
426425177
MDR Report Key11182489
MDR Text Key240671575
Report Number8010047-2021-01507
Device Sequence Number1
Product Code ITX
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
K001203
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Other
Type of Report Initial
Report Date 01/16/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received01/15/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator No Information
Device Model NumberUM-G20-29R-3
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/09/2020
Was the Report Sent to FDA? No
Date Manufacturer Received08/25/2020
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured04/15/2020
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-