• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION RESONATE X4 CRT-D; IMPLANTABLE DEVICE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION RESONATE X4 CRT-D; IMPLANTABLE DEVICE Back to Search Results
Model Number G447
Device Problems High impedance (1291); Noise, Audible (3273)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 12/17/2020
Event Type  malfunction  
Manufacturer Narrative
No further information is available at this time.If information is provided in the future, a supplemental report will be issued.
 
Event Description
It was reported that this cardiac resynchronization therapy defibrillator (crt-d) emitted beeping tones.Data was reviewed by technical services (ts) in which the cause was attributed to a high out of range shock impedance measurement.Further device evaluation is expected at a future date.This device remains in service.No adverse patient effects were reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
RESONATE X4 CRT-D
Type of Device
IMPLANTABLE DEVICE
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
4100 hamline avenue north
saint paul MN 55112
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
cashel road
clonmel
EI  
Manufacturer Contact
timothy degroot
4100 hamline avenue north
saint paul, MN 55112
6515826168
MDR Report Key11182961
MDR Text Key227558991
Report Number2124215-2021-00513
Device Sequence Number1
Product Code NIK
Combination Product (y/n)N
Reporter Country CodeUK
PMA/PMN Number
P010012/S436
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type consumer
Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date 01/17/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Expiration Date04/23/2022
Device Model NumberG447
Device Catalogue NumberG447
Device Lot Number237346
Was Device Available for Evaluation? No
Initial Date Manufacturer Received 12/17/2020
Initial Date FDA Received01/17/2021
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured05/05/2020
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-