MAUDE Adverse Event Report: MEDTRONIC HEART VALVES DIVISION ENVEO PRO DELIVERY SYSTEM; AORTIC VALVE, PROSTHESIS, PERCUTANEOUSLY DELIVERED

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC HEART VALVES DIVISION ENVEO PRO DELIVERY SYSTEM; AORTIC VALVE, PROSTHESIS, PERCUTANEOUSLY DELIVERED Back to Search Results
Model Number ENVPRO-16
Device Problems Difficult or Delayed Positioning (1157); Detachment of Device or Device Component (2907)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 12/28/2020
Event Type  malfunction  
Manufacturer Narrative
Product analysis: the device was not returned, therefore no product analysis can be performed.Conclusion: without return of the product, no definitive conclusions could be drawn regarding the clinical observation.If information is provided in the future, a supplemental report will be issued.
 
Event Description
Medtronic received information that during the implant of this transcatheter bioprosthetic valve, the valve was 2/3 deployed when it was decided to recapture the valve.When the blue handle of the delivery catheter system (dcs) started to turn towards the close, the handle began to split into two.The handle could not be closed again so the physician broke the remaining handle and released the valve the way it was.The valve position was adequate.No adverse patient effects were reported.
 
Manufacturer Narrative
Product analysis:upon receipt at medtronic's quality laboratory, the device was received with the handle components detached from the delivery catheter system (dcs).The device was received with the capsule over captured.The trigger moved to fully advanced and retracted positions and locked in place when released.The tip-retrieval mechanism appeared intact.Delamination was observed over the nitinol reinforcing frame along the proximal end of the capsule.The inner member shaft and spindle hub appeared intact with no evidence of damage.Tab 3 was observed to be hinged inward.Tab 4 was observed to have detached from the male actuator component.The device was returned with the end cap/screw gear snap fit connected.The reported event for handle separation could be confirmed in the analysis.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Manufacturer Narrative
Conclusion: the device history record was reviewed and showed that this product met all manufacturing specifications for product released for distribution.No issues were identified that would have impacted this event.Recapturing is a feature of the evolut system that allows for additional attempts at accurately positioning the valve.Various factors can affect valve positioning including patient anatomy or physician technique.There is no information to suggest a device quality deficiency that may have caused or contributed to this event and the root cause of the positioning difficulty could not be conclusively determined with the available evidence and a relationship to the dcs cannot be established.Delamination was observed over the nitinol reinforcing frame along the mid-section to the proximal end of the capsule, which typically occurs when the capsule is subjected to a bending force potentially after tracking through tortuous anatomy.Actuator separation is typically associated with broken or damaged snap fit tabs, either one tab or both, which hold the handle together.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Manufacturer Narrative
Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ENVEO PRO DELIVERY SYSTEM
Type of Device
AORTIC VALVE, PROSTHESIS, PERCUTANEOUSLY DELIVERED
Manufacturer (Section D)
MEDTRONIC HEART VALVES DIVISION
1851 e deere ave
santa ana CA 92705
Manufacturer (Section G)
MEDTRONIC HEART VALVES DIVISION
1851 e deere ave
santa ana CA 92705
Manufacturer Contact
toni o'doherty
parkmore business park west
galway 
EI  
091708734
MDR Report Key11194773
MDR Text Key227573029
Report Number2025587-2021-00201
Device Sequence Number1
Product Code NPT
Combination Product (y/n)N
Reporter Country CodePO
PMA/PMN Number
P130021
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative
Reporter Occupation Physician
Remedial Action Recall
Type of Report Initial,Followup,Followup,Followup
Report Date 10/28/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date04/16/2022
Device Model NumberENVPRO-16
Device Catalogue NumberENVPRO-16
Device Lot Number0010217153
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/01/2021
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 12/28/2020
Initial Date FDA Received01/19/2021
Supplement Dates Manufacturer Received02/09/2021
02/10/2021
03/20/2022
Supplement Dates FDA Received02/10/2021
02/11/2021
10/28/2022
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured04/16/2020
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Removal/Correction Number2025587-05-28-2021-001-R
Patient Sequence Number1
Patient Age70 YR
Patient SexMale
Patient Weight68 KG
-
-