  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SIEMENS HEALTHCARE GMBH- AT ARTIS ZEE FLOOR; INTERVENTIONAL FLUOROSCOPIC X-RAY SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SIEMENS HEALTHCARE GMBH- AT ARTIS ZEE FLOOR; INTERVENTIONAL FLUOROSCOPIC X-RAY SYSTEM Back to Search Results
Model Number 10094135
Device Problem Mechanical Problem (1384)
Patient Problem Cardiac Arrest (1762)
Event Date 12/27/2020
Event Type  Death  
Manufacturer Narrative
Siemens is conducting a thorough investigation of the reported events.As this event is under investigation, a root cause has not yet been determined.A supplement report will be filed upon completion of the investigation.
 
Event Description
It was reported to siemens that an adverse event occurred while operating the artis zee floor system.During an interventional procedure, the user reported that the stand would not move at all and the system went down.It was later reported that the patient was admitted in a critical condition and passed away as the result of a pulmonary embolism.At this time there is no causal relationship between the patient condition and the system malfunction.Siemens has requested additional information in order to conduct an investigation of the reported event.
 
Manufacturer Narrative
Siemens has completed an investigation of the reported event.The root cause was determined to be a hardware error.The cause of the fault was a defective potentiometer.The potentiometer was replaced on site and the system again functioned as specified.The problem described has not been reported again.The spare parts consumption of the affected component was checked and a possible general error that would require corrective actions of the installed base could not be determined by the investigation.According to the information available, the patient was already in a critical state when he arrived at the hospital.After consultation with the physicians, no causal connection between the above-mentioned malfunction and the patient's passing can be identified.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ARTIS ZEE FLOOR
Type of Device
INTERVENTIONAL FLUOROSCOPIC X-RAY SYSTEM
Manufacturer (Section D)
SIEMENS HEALTHCARE GMBH- AT
siemensstrasse 1
forchheim, germany 91301
GM  91301
MDR Report Key11200699
MDR Text Key227723273
Report Number3004977335-2021-62451
Device Sequence Number1
Product Code OWB
UDI-Device Identifier04056869010045
UDI-Public04056869010045
Combination Product (y/n)N
PMA/PMN Number
K181407
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Remedial Action Repair
Type of Report Initial,Followup
Report Date 01/20/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received01/20/2021
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Model Number10094135
Was Device Available for Evaluation? Yes
Date Manufacturer Received02/23/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Outcome(s) Death;
-
-