Model Number N/A |
Device Problems
Loss of Power (1475); Output below Specifications (3004)
|
Patient Problems
No Consequences Or Impact To Patient (2199); No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 01/08/2021 |
Event Type
malfunction
|
Manufacturer Narrative
|
This event has been recorded by zimmer biomet under (b)(4).Once the investigation is completed, a follow-up/final report will be submitted.
|
|
Event Description
|
It was reported that the device lost power during surgery.No adverse events were reported as a result of this event.
|
|
Manufacturer Narrative
|
This event has been recorded by zimmer biomet under (b)(4).This medwatch is being filed to relay additional information.Review of the most recent repair record identified no repairs related to the reported event.Device is used for treatment.It was noted the device has not been returned regularly for recommended annual pm.A definitive root cause cannot be determined as no problem was found related to the reported event.The event cannot be confirmed.
|
|
Event Description
|
There is no additional information.
|
|