According to the reporter, when the surgeon was trying to ligate the cystic duct during a laparoscopic cholecystectomy, the surgeon was able to squeeze the handle but there were issues experienced with the loading or firing of the clip.It was also stated that the clip applier contained less than 16 clips and the jaws of the device was not able to close when the surgeon tried to squeeze the applier.Another device was used.There was no patient injury.
|
H3 evaluation summary: medtronic conducted an investigation based upon all information received.The device was available for evaluation.Visual inspection found a jaw leg that was out of position.The cam on the jaw leg was not aligned with the driver.It was reported that the clips did not load properly into the jaws as expected.The reported issue was confirmed.The product analysis noted evidence that the device was not used as intended.This issue can occur if one of the jaws legs gets forcefully pulled outward away from the center line of the shaft.The noted jaw cam disengagement impedes functionality of the jaws, possibly resulting in clip malformation or difficulty removing the instrument from a trocar.It was also reported that the instrument had less than the expected amount of clips.The reported issue could not be confirmed.The product analysis noted evidence that the device was not used as intended.The manufacturing records for each device are thoroughly reviewed prior to release to ensure that it meets all medtronic quality specifications.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|