SMITH & NEPHEW, INC. LGN 6 DEG VAL COL 6MM OFF S3-8; PROSTHESIS, KNEE, FEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, METAL/POLYMER
|
Back to Search Results |
|
Model Number 71434407 |
Device Problem
Mechanical Jam (2983)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 01/04/2021 |
Event Type
malfunction
|
Manufacturer Narrative
|
(b)(4).
|
|
Event Description
|
It was reported that, the lgn 6 deg val col 6mm off s3-8 is jammed and unable to rotate preventing the surgeon to dial in the correct rotation.It is unknown how the procedure was completed.There was not delay.No patient injury or other complications were reported.
|
|
Manufacturer Narrative
|
The device, intended for use in treatment was returned for evaluation: a visual inspection was performed and confirmed the device appear to be jammed enabling it to turn.The manufactured date for this device is 2011.A review of the manufacturing records did not reveal a manufacturing abnormality that could have caused or contributed to the reported incident.A complaint history review found related failures; this failure mode will be monitored for future complaints for any necessary corrective actions.This device is a reusable instrument that can be exposed to numerous surgeries; damage from repeated use can occur.Damage from prolonged use, misuse or rough handling are likely probable causes of the reported event.We recommend that all reusable instruments be routinely inspected for wear and damage and replaced as necessary.At this time, we have no reason to suspect that the product failed to meet any product specifications at the time of manufacture.Based on this investigation, the need for corrective action is not indicated.No further investigation is warranted for this complaint; however, we will continue to monitor for future complaints and investigate as necessary.Should additional information be received, the complaint will be reopened.
|
|
Search Alerts/Recalls
|
|
|